System Operation
This chapter details the “Cool down” and “Warm up” procedures for the cryogenic components of the Opus One nanowire system.
Cool Down Procedure
For Opus One Integrated and Opus One Portable Units
For systems with one-button cooldown, the Opus One is able to autonomously handle normal operating conditions and has built-in safeguards for equipment or power failure. To begin system cooldown, follow the steps below.
Switch the electronics on. The cooldown button will turn green.
Press the button and the system will begin cooling. The button will turn from green to magenta, and the vacuum pump will turn on after a 1 minute delay.
Once sufficient time has elapsed, the button will turn to cyan when the vacuum pump relay is switched off, and will turn to dark blue when the system is fully cold. You may confirm that the temperature reading of the Stage 1 thermometer indicates the temperature has started to decrease via the Opus One controller software.
For Opus One Standard Units
The procedures in this section should be accomplished for first-time operation and repeated any time that the system has been warmed such that both stages are above 50 Kelvin, for example in case of power failure, cooling water loss, or intentional system shutdown.
The cool down procedure steps should be completed in the following order when the system is fully warm:
The cool down procedure steps should be completed in the following order when the system is fully warm:
- Evacuate the Opus One unit according to the following procedure.
- Ensure the green vacuum valve (labeled "Valve") on the rear of the Opus One unit is closed by gently applying clockwise torque. The valve should not open under light counter-clockwise torque. Do not overtighten.
- Remove the NW25 cap and o-ring from the vacuum port (labeled "Pump/Vent") by loosening the clamp screw and removing the clamp.
- Inspect the o-ring to ensure it is dust-free and has no cracks or flaws. If so, replace the o-ring before proceeding.
- Using the o-ring and clamp, connect the vacuum pump via an appropriate NW25 vacuum tube (not provided) to the vacuum port and hand tighten the clamp securely.
- Follow the vacuum pump manufacturer's recommended procedure for starting the pump.
- Slowly open the vacuum valve on the Opus One unit. The valve is fully open once rotated counterclockwise approximately two (2) full turns. The valve should remain open and the pump on until the system has reached 50 K.
- Ensure the compressor is ready for use, including turning on any required cooling water or electrical breakers.
- Turn on the electrical power to the compressor. The hum of the compressor and the 1 cycle-per-second "hiss" of the Opus One unit should begin immediately. If the 1 cycle-per-second sound from the Opus One is not heard, turn off the compressor and confirm that the helium line and electrical connections between the Opus One unit and the compressor are correct, as per Section 2.2.3.
- Confirm that the temperature reading of the Stage 1 thermometer indicates the temperature has started to decrease.
- After the temperature of the Stage 1 thermometer drops below approximately 50 K, close the vent valve by turning the green valve knob clockwise until (gently) finger tight. Do not overtighten.
Ensure the compressor is ready for use, including turning on any required cooling water or electrical breakers.
Turn on the electrical power to the compressor. The hum of the compressor and the 1 cycle-per-second "hiss" of the Opus One unit should begin immediately. If the 1 cycle-per-second sound from the Opus One is not heard, turn off the compressor and confirm that the helium line and electrical connections between the Opus One unit and the compressor are correct, as per Section 2.2.3.
Confirm that the temperature reading of the Stage 1 thermometer indicates the temperature has started to decrease.
After the temperature of the Stage 1 thermometer drops below approximately 50 K, close the vent valve by turning the green valve knob clockwise until (gently) finger tight. Do not overtighten.
Once the vacuum valve is closed you may disconnect the pumping line from the Opus One and turn off the pump, according to the process recommended by the pump manufacturer.
Seal the NW25 Pump/Vent port using the NW25 o-ring, cap, and clamp.
Wait until the temperature of the Stage 2 thermometer indicates the system is at its "base" temperature (typically below 2.5 K). The cool-down takes approximately three to four (3-4) hours from the time the compressor is turned on.
The Opus One devices are now ready to be used. Refer to Section 5.1.
Warm Up Procedure
For Opus One Integrated and Opus One Portable Units
To warm the system, press the lit button on the front of the Opus One. The light will turn from its current status color to green, the compressor will turn off, and the system will begin warming.
For Opus One Standard Units
To warm the system, turn off the compressor. If using a water-cooled compressor, the water cooling can be turned off at this time as well.